The VT8000 covers ICU ventilation in a single device: invasive, noninvasive, and high-flow, for adults and children, and for newborns with the optional module. It comes with a complete set of lung-protective tools, and its interface is built so your team can work with it confidently from the start.
All established invasive and noninvasive modes in one device, from VCV and PRVC to BiVENT and APRV.
P/V maneuver, lung recruitment, C20/C, and SI help reduce ventilator-induced lung injury (VILI).
High-flow nasal cannula for noninvasive respiratory support.
A dedicated mode for ventilation during resuscitation.
Structured weaning with SBT, NIF, P0.1, RSBI, and WOB.
Real-time display of resistance, compliance, and breathing status.

Clearly laid out, with no hidden menus. Your team gets up to speed quickly and can stay focused on the patient.

All the relevant curves and parameters can be shown together on one screen, arranged to your preferences and the clinical situation.

The "Dynamic Lung" view shows resistance, compliance, and the patient's breathing and ventilation status graphically in real time. It gives you a quick picture of lung status and supports your treatment decision.
Continuous pulse oximetry with Masimo rainbow SET for reliable readings during motion and low perfusion.
Mainstream capnography with a rise time (T90) under 90 ms and warm-up under 10 s.
Ventilation for newborns with a proximal flow sensor and a minimum tidal volume of 2 ml.
| Patient groups | Adults, children, newborns (neonatal optional) |
|---|---|
| Tidal volume | Adults 50–4000 ml · Children 20–300 ml · Neonatal from 2 ml (opt.) |
| Ventilation modes | VCV, PCV, PRVC, SIMV, SPONT, VSV, BiVENT, APRV, CPRV, NIV-CPAP, NIV-PCV, nCPAP, HFNC |
| Max. working pressure | 100 cmH₂O (pressure limit 125 cmH₂O) |
| Trigger | Flow and pressure |
| Display | 15.6″ TFT, 1920×1080, capacitive touchscreen, multi-angle view |
| Weight / dimensions | 25 kg (without trolley) · 290 × 236 × 398 mm |
| Power supply | 100–240 VAC / 24 VDC, 50/60 Hz |
| Battery runtime | ≥ 2 h (VCV mode), charge time 4 h |
| Applied standards | ISO 80601-2-12 (ventilation), -2-55 (gas monitor), -2-90 (high-flow), IEC 60601-1 |
| SpO₂ | Masimo rainbow SET |
| CO₂ (IRMA mainstream) | Range 0–15 vol % ±(0.2 vol % + 2 %) · rise time (T90) ≤ 90 ms · warm-up < 10 s |
| Neonatal | Proximal flow sensor, tidal volume from 2 ml |
| Additional | Ptp/Pes, NIBP, nebulizer |
Adults, children and newborns. Neonatal ventilation is available via the optional neonatal module.
Conventional invasive and non-invasive modes as well as an HFNC mode and a CPRV mode to support ventilation during resuscitation.
Through Lung Protective Ventilation Strategies (LPVS) – P/V manoeuvre, lung recruitment, the C20/C index and Sustained Inflation – to reduce ventilator-induced lung injury (VILI).
Optionally an SpO₂ module (Masimo rainbow SET), a CO₂ module (IRMA mainstream) and a neonatal module.
The VT8000 has a 15.6″ TFT touchscreen with a multi-angle view.